CL-10 Plus – User Manual
57 / 75
ECPUS vers 0.2 eng
5. Maintenance and Preservation
5.1 Weekly
Maintenance
Run the following procedure:
→
at least 12 hours before the working session, on a weekly base, or
→
when the Blank value, during calibration procedure, is beyond the limits accepted by the
method, or
→
after changing any fluidic component (pump tubing’s, internal tubing’s, parts of mixing
chamber, etc).
1. Start the CL10 Plus program.
2. Replace the wash solution with REGENERATING SOLUTION.
3. Check that in starter position there is a vial containing distilled water.
4. Run 1
Prime enzyme
and 2
Clean
.
5. Wait until completion of cycles and 5 minutes more.
6. Replace the REGENERATING SOLUTION with reconstituted wash solution.
7. Run 1
Prime enzyme
and 2
Clean
.
8. Leave the instrument in
Rest
mode (see § 4.6.2).
5.2 Routine
Maintenance
Run the following procedure:
→
regularly every 3 months
→
at least 1 day before working session, after extended period halt
→
when the ‘
enzyme Contamination’ occurs (see. § 5.4 )
1. Start the CL10 Plus program.
2. Replace the wash solution with STRONG REGENERATING SOLUTION.
3. Check that in starter position there is a vial containing distilled water.
4. Run 1
Prime
enzyme
and 2
Clean
.
5. Wait until completion of cycles and 3 minutes more.
6. Replace the STRONG REGENERATING SOLUTION with distilled water.
7. Run 1
Prime
enzyme
and 2
Clean
.
8. Replace the distilled water with reconstituted wash solution.
9. Run 2
Clean
cycles.
10. Leave the instrument in
Rest
mode (see § 4.6.2).