1)
OFFSET
.
First step: run of an analysis cycle without injecting any milk in the mixing chamber and
obtaining:
D1: the two electrodes pumps (1 and 2) turn and take the mixer content, only buffer; when
the electrodes contain the same solution, their pH difference remains more or less the
same, so for 5 seconds it is approx. 20
∆
mpH.
D2: the enzyme pump (6) loads the urease enzyme into the mixer and only the electrode 2
takes the solution urease, while the electrode 1 contains only buffer; the urease
may have a little difference of pH compared to the buffer, due to the enzyme contribution
to the reading.
In the graph the reaction is as follows:
During the D2 there is no reaction, but only a measure of the different “acidity” due to the
enzyme.
So, if you remember the graph of our milk sample, not all the 10
∆
mpH are caused by the
urea hydrolysis, 1
∆
mpH is in fact due to the enzyme contribution:
∆
mpH
t (s)
20
5
30
21
(DL1)
D1 time
D2 time
∆
mpH
t (s)
20
5
30
30
(DL1)
21
D1 time D2
time