CL-10 Plus – User Manual
67 / 75
ECPUS vers 0.2 eng
1. Replace the reconstituted wash solution with R1 buffer of Electrode test kit.
2. Disconnect the starter tube (B, Figure 2) from the steel capillary (A) and insert it in a vial
containing distilled water (the same where also the other end C, is placed)
3. Follow the instructions of package insert and fill in the form:
Report Electrodes performance CL-10 Plus
(available upon request and included in the kit
GEN603).
A
B
C
FIGURE 2:
STARTER line
4. If necessary, print the graph of the test (use the
Print Graph
icon
)
, after placing the limits
at +/-100 from the symmetry line of the graph.