CL-10 Plus – User Manual
53 / 75
ECPUS vers 0.2 eng
4.6.4 Calibrate
Temperature.
This command is only used by technical personnel.
CL-10 Plus contains in the hardware all information for the right temperature during the
working session.
To check that this setting is not idle:
From
Explorer
Enter CL10 folder and open the text file (.txt): C:\cl10\cl10.ini:
cl10.ini
Under the Header [
OPERATIONS
], verify that TemperatureOffset = 0. Digit zero if a
different value is displayed. From
File
menu select
Save
and close the window.
Once this is made, it is possible to change the PC without affecting the temperature offset;
in fact this parameter remains completely independent from the PC used and it remains
stored in the instrument hardware.
4.6.5
Communication parameters.
This window is used if it is required to configure and select the serial port (RS 232) used
by CL10 Plus.
The preferential channel is COM1, the alternative one is COM2, in case COM 1 is
engaged in other jobs. COM3 and COM4 are not enabled.
Concerning Baud Rate (9600), Data bits (8 bit), Stop bits (1 stop), Parity (None), the
respective values must be as described in the picture; in any case the same as the
interface values used by the PC.
4.6.6
Print Technical Report.
This command is used when required to check the instrument performance of a working
session. The function is only used by technical personnel from , unless the CL-10 operator
is addressed to cooperate.
The printing procedure is the same as
Worklist
command (see § 4.5.1); although the
printout has a different structure and a number of additional information (D1, D2,
∆
mpH)
compared to
Print Worklist
, allowing technical specialised personnel to detect possible
causes of faults.