CL-10 Plus – User Manual
64 / 75
ECPUS vers 0.2 eng
Close the panel, fix the pumps into their seats and switch the instrument on.
If the problem persists, ask for technical assistance.
6.1.3
The analytical unit does not accept commands from the program.
•
The “CL10 Protocol Error” following box appears on the screen.
•
Check that the instrument is on
•
Check that the CL-10TBL_BUILDER program is off (if not, turn it off)
•
Check the connections between the serial port RS 232 (COM 1 in the PC) and the
analyser; if the instrument is connected to the wrong serial port (COM 2), connect it to
the serial port COM 1, or select the COM 2 port from
Communication Parameters
in
SERVICE
menu (see following windows):
•
If the problem persists, ask for technical assistance.