CL-10 Plus – User Manual
10 / 75
ECPUS vers 0.2 eng
2.3.6 Internal
&
peristaltic pump tubing’s
Internal tubing’s are 8 pcs (SPP311); normally they are not considered as
consumables, unless accidentally broken or damaged.
Peristaltic tubing’s, or pump tubing’s, are consumables and as such, they must be
checked before installation and inspected every three months.
It is recommended the replacement every 3000 cycles, that means in average,
3-5 months of work.
Description of tubing types, and destination:
Pos
Description & Packaging
Code
1
Internal Tubing’s set
8x
SPP 311
2
Peristaltic Pump Tubing’s Set
4x
SPP 312
3
Enzyme Starter tubing
only mono starter
1x
SPP 313
4
Additional Tubing: starter
Wine Combi
1+1
SPP 314
5
Additional Tubing: specific
Wine pH/Acidity
1+1
SPP 315
6
Additional Tubing’s: double starter
Bio G6PD/6PGD
1+1
SPP 316
SPP 311 : product for all configurations and all applications
SPP 312:
product for all configurations and all applications
(pump tubing’s 1-2-3-4, Fig 2 § 2.3.5)
SPP 313: product for all kits other than WCP678-682, BCP 592-596-690 (pump
tubing 6, Fig 2 § 2.3.5)
SPP 314:
product for WCP678, only (pump tubing’s 5-6, Fig 2 § 2.3.5)
SPP 315:
product for WCP682, only (pump tubing’s 5-6, Fig 2 § 2.3.5)
SPP 316:
product for BCP 592-596-690, only
(pump tubing’s 5-6, Fig 2 § 2.3.5)