CL-10 Plus – User Manual
5 / 75
ECPUS vers 0.2 eng
2. Installation
The CL10 Plus instrument consists of a thermo stable measuring block which contains two
capillary glass electrodes, a mixing chamber, two groups of peristaltic pumps, a differential
amplifier and a microprocessor which calculates the analyte concentration or enzyme activity
in the sample, and controls all the functions of the instrument.
CL-10 Plus is delivered with the fluidics empty (free from liquid) and dry and with the
peristaltic pumps unhooked (see § 5.5 figure; the detail of the pump tubing’s).
The procedure for the electrodes reactivation and stabilisation takes from 2 to 24 hours (see
§ 2.5.3).
2.1 . Unpacking and checking.
Check the list of content:
1.
Instrument CL10 Plus
2.
power 230 VAC (110 VAC) - 12 VDC
3.
power cable
4.
serial cable CL10-PC
5.
Plastic bottles ( Kartel type) for the buffer and/or standby solution and for the waste
(2x250 ml)
6.
Set pump tubing’s (Art. No. SSP112 or SPP113)
7.
CL-10 Plus Software on CD-ROM and relative Licence
8.
CL-10 Plus Operator Manual
9.
CL10 Plus Testing Certificate
Instrument Serial Number
:
The serial number is located on the back of the instrument; it is the same number
written on the CL-10 Plus Testing Certificate .
2.2 . PC system Configuration.
•
Pentium PC at 175 MHz or superior
•
Microsoft
Windows
95/98/2000/Me/XP
•
RAM 32 MB or superior
•
6 MB available on the Hard Disk
•
floppy disk 3,5”
•
Monitor Super VGA with 256 colours
•
Video card with 800x 600 pixel
•
CD-ROM 4x or superior
•
COM1 RS232 9 pin or USB* port
•
If USB port is used USB-RS232 converter is required.
(recommended: Targus Replicator type. PA070)