8 Appendices
Rod source shields LB 8300
46
64217BA2 Rev. 00, 10/2017
Checking the functions of the locking mechanism
Requirement
met
not met
It must be possible to move the locking
mechanism completely back and forth
between the two positions (OPEN /
CLOSED).
The measuring signal of the
measurement system must
approximately correspond to the
maximum value in the CLOSED
position.
If alternatively a dose rate
measurement device is used, the dose
rate in the CLOSED position must be
clearly lower than in the OPEN
position.
Only with manual actuation:
It must be possible to turn the locking
cylinder using one hand and the lever
rod.
Only with pneumatic actuation:
The air pressure required for moving
the locking mechanism may not
exceed the maximum permissible air
pressure (see technical data of the
pneumatic actuators)
Name of inspector:
Date:
Signature: