Rod source shields LB 8300
4 Installation
64217BA2 Rev. 00, 10/2017
21
4.6
Installation of the shield
To avoid unnecessary exposure to radiation, install the shield as final system
component.
The shield is placed on a mounting base at the measuring point and fixed. The
shield on the container is secured against tilting by means of a tipping protection.
j
WARNING
Risk of injury due to tilting shield
There is a risk of injury to persons when working on non-fixed shields.
Always secure upright shields against falling by fasten lifting means at the
ring nuts.
IMPORTANT
The specification of the mounting base and the tilt safety device are
determined during the project planning and defined by drawings, sketches or
written instructions.
These instructions must be observed strictly during assembly.
Tip
If the shield contains a radioactive source, the measurement of the
background count rate of the detector should take place before the shield is
installed.