Level Detector Error Codes (Continued)
Table 4.2
Sensor
Problem
Solution
QJ71GF ERR
The PLC module (QJ71GF)
used for the
communication between
the PLC and LC
malfunctioned.
1. Reset the PLC.
2. If the reset does not recover from the error, QJ71GF
module for the PLC can be defective.
REG RST
PLC register settings have
been reset. The settings
may have reverted to their
default values.
Inspect and adjust the PLC register settings as required.
SL001 ON
SL001 OFF
SL002 ON
SL002 OFF
SL003 ON
SL003 OFF
SL005 ON
SL005 OFF
Stopper error. The stopper
is not working correctly.
ON error: The cylinder does
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
1. Confirm that a carrier is not jammed or stuck.
2. Confirm that the air compressor is on and the auto switch
is working correctly.
3. Confirm that the compressed air supply is within the
normal pressure range.
4. Select
Restart
on the control panel to resume
operation.
SL011A ON
SL011B ON
SL011 OFF
Error at the Y-axis cylinder
on the carrier transfer arm.
ON error: The cylinder does
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
1. Confirm that the air compressor is on and the auto switch
is working correctly.
2. Confirm that the compressed air supply is within the
normal pressure range.
3. Confirm that nothing interferes with the movement of the
loading arm.
4. Inspect the transfer arm for misalignment.
SL013 ON
SL013 OFF
Error at the Z-axis cylinder
on the carrier transfer arm.
ON error: The cylinder does
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
1. Confirm that the air compressor is on and the auto switch
is working correctly.
2. Confirm that the compressed air supply is within the
normal pressure range.
3. Confirm that nothing interferes with the movement of the
loading arm.
4. Inspect the transfer arm for misalignment.
Troubleshooting
Error Code Tables
4
B31449AE
4-5
Содержание Power Express Aliquot
Страница 6: ...Warranty and Service vi B31449AE...
Страница 22: ...Introduction Scope of this Manual xxii B31449AE...
Страница 30: ...Power Express Aliquot Module Description Components 1 8 B31449AE...
Страница 110: ...Figure 4 3 Tube Supply Sensor Diagram 4 50 B31449AE Troubleshooting...
Страница 111: ...Figure 4 4 Label Supply Sensor Diagram B31449AE 4 51 Troubleshooting...
Страница 112: ...Aliquot Unit Figure 4 5 Overview Sensor Diagram 4 52 B31449AE Troubleshooting...
Страница 114: ...Figure 4 7 Tip Supply Sensor Diagram 4 54 B31449AE Troubleshooting...
Страница 117: ...Figure 4 10 Optional Track Sensor Diagram B31449AE 4 57 Troubleshooting...
Страница 118: ...Figure 4 11 Merge Lane Sensor Diagram 4 58 B31449AE Troubleshooting...
Страница 119: ...APPENDIX A Related Documents Related Documents General System Operation IFU PN B31448 B31449AE A 1...
Страница 120: ...Related Documents Related Documents A 2 B31449AE...
Страница 121: ......
Страница 122: ...www beckmancoulter com 2018 Beckman Coulter Inc All Rights Reserved...