T Lane Error Codes (Continued)
Table 4.6
Sensor
Problem
Solution
CCIE ERR
Established communication
between PLC and LC has
been disrupted.
1. Confirm the cable connection between LC and PLC, located
at the upper stream side of the Error unit.
2. Confirm all the PLCs within the line are powered on.
3. When the communication between PLC and LC has been
re-established, an error is recovered.
4. Select
Restart
on the control panel to retry or use
Function code 01 (Lane through (Bypass)) to continue.
CCIE SND
Failed to send a message to
the line controller.
Confirm that the cables between individual PLCs and the Line
Control are connected correctly.
OP CALL
PLC hardware error. A PLC
battery may have run out.
Replace the battery if the battery indicator on the front of the
PLC box is blinking.
PS065 PAS
Passage sensor error. A
passage sensor does not
detect a carrier within a
predetermined timeout
period.
1. Confirm that a carrier is not jammed or stuck and that the
belt moves.
2. Select
Restart
on the control panel to resume
operation.
QJ71GF ERR
The PLC module (QJ71GF)
used for the
communication between
the PLC and LC
malfunctioned.
1. Reset the PLC
2. If the reset does not recover from the error, QJ71GF
module for the PLC can be defective.
RF01 INI
RF02 INI
RF03 INI
RFID initialization error.
1. Confirm that the RFID cable is connected correctly.
2. Reboot the module.
3. Place the module in
Run
mode to resume operation.
REG RST
PLC register settings have
been reset. They may have
reverted to their default
values.
Inspect and adjust the PLC register settings as required.
SL001 ON
SL001 OFF
SL002 ON
SL002 OFF
SL003 ON
SL003 OFF
SL004 ON
SL004 OFF
Stopper error. The stopper
is not working correctly.
ON error: The cylinder does
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
1. Confirm that a carrier is not jammed or stuck.
2. Confirm that the air compressor is on and the auto switch
is working correctly.
3. Confirm that the compressed air supply is within the
normal pressure range.
4. Select
Restart
on the control panel to resume
operation.
Troubleshooting
Error Code Tables
4-40
B31449AE
Содержание Power Express Aliquot
Страница 6: ...Warranty and Service vi B31449AE...
Страница 22: ...Introduction Scope of this Manual xxii B31449AE...
Страница 30: ...Power Express Aliquot Module Description Components 1 8 B31449AE...
Страница 110: ...Figure 4 3 Tube Supply Sensor Diagram 4 50 B31449AE Troubleshooting...
Страница 111: ...Figure 4 4 Label Supply Sensor Diagram B31449AE 4 51 Troubleshooting...
Страница 112: ...Aliquot Unit Figure 4 5 Overview Sensor Diagram 4 52 B31449AE Troubleshooting...
Страница 114: ...Figure 4 7 Tip Supply Sensor Diagram 4 54 B31449AE Troubleshooting...
Страница 117: ...Figure 4 10 Optional Track Sensor Diagram B31449AE 4 57 Troubleshooting...
Страница 118: ...Figure 4 11 Merge Lane Sensor Diagram 4 58 B31449AE Troubleshooting...
Страница 119: ...APPENDIX A Related Documents Related Documents General System Operation IFU PN B31448 B31449AE A 1...
Страница 120: ...Related Documents Related Documents A 2 B31449AE...
Страница 121: ......
Страница 122: ...www beckmancoulter com 2018 Beckman Coulter Inc All Rights Reserved...