If...
Then...
samples are
completed,
Open the connection unit door, retrieve the samples. Transfer these
tubes to the sample tube carriers that are waiting to be unloaded at the
Connection Module (SN031). Place the Connection Unit in Mode
96
(Recovery) to release the completed samples. Close the connection unit
door.
samples are not
completed,
Open the connection unit door, retrieve the samples. Transfer these
tubes to an offline instrument rack and load the samples as a STAT rack.
When completed, manually unload the samples and place the tubes into
the sample tube carriers that are waiting to be unloaded at the
Connection Module (SN031). Close the connection unit door.
e. On the analyzer connection module control panel, Select
Enter
and
Exit
.
16
At the AU5800 and AU5800XL Rack Builder modules:
a. If a stopper is jamming the sample tube carrier, push it forward to release the
carrier.
b. If a tube is in the loading arm gripper, remove the tube and place back into the
empty sample carrier at the barcode reader.
c. If tubes are in the loading racks remove the tubes and set aside.
d. If tubes are in the unloading arm grippers remove the tubes and place the tubes
back into the rack at the Unloading position. Re-adjust any other tubes in the
Unload rack so that the tubes start at position 1 and have no empty spaces.
e. Block any tubes behind RF01 position. Select
Auto
and
OK
to place the
module back into operation. Allow the system to load any tubes at RF01 into the
loading rack. The system blocks remaining tubes behind RF01.
f. Place the tubes that you removed in the shuttle carrier cups at the orientation
station one by one.
g. The system loads tubes in the orientation stations into the loading rack.
h. Repeat steps f and g until all tubes are reloaded.
i. When completed, remove the blockage at RF01.
17
At the DxH Rack Builder modules:
a. If a stopper is jamming the sample tube carrier, push it forward to release the
carrier.
b. If a tube is in the loading arm gripper, remove the tube and place back into the
empty sample carrier at the barcode reader.
c. If tubes are in the loading racks remove the tubes and set aside. If there are any
loading racks in the upper transfer shuttle, remove them and collect the tubes from
the racks.
d. If a tube is in the unloading arm gripper remove the tubes and place the tubes back
into the rack at the Unloading position. Re-adjust any other tubes in the Unload rack
so that the tubes start at position 1 and have no empty spaces. If there are any racks
in the upper unload shuttle area, remove the racks and set them aside.
Error Recovery Procedures
Power Express Aliquot Error Recovery Procedures
3
B31449AE
3-11
Содержание Power Express Aliquot
Страница 6: ...Warranty and Service vi B31449AE...
Страница 22: ...Introduction Scope of this Manual xxii B31449AE...
Страница 30: ...Power Express Aliquot Module Description Components 1 8 B31449AE...
Страница 110: ...Figure 4 3 Tube Supply Sensor Diagram 4 50 B31449AE Troubleshooting...
Страница 111: ...Figure 4 4 Label Supply Sensor Diagram B31449AE 4 51 Troubleshooting...
Страница 112: ...Aliquot Unit Figure 4 5 Overview Sensor Diagram 4 52 B31449AE Troubleshooting...
Страница 114: ...Figure 4 7 Tip Supply Sensor Diagram 4 54 B31449AE Troubleshooting...
Страница 117: ...Figure 4 10 Optional Track Sensor Diagram B31449AE 4 57 Troubleshooting...
Страница 118: ...Figure 4 11 Merge Lane Sensor Diagram 4 58 B31449AE Troubleshooting...
Страница 119: ...APPENDIX A Related Documents Related Documents General System Operation IFU PN B31448 B31449AE A 1...
Страница 120: ...Related Documents Related Documents A 2 B31449AE...
Страница 121: ......
Страница 122: ...www beckmancoulter com 2018 Beckman Coulter Inc All Rights Reserved...