Error Recovery General Information
When an error condition occurs, the system generates an error message, sounds an alarm
and flashes a warning beacon. Error messages are shown on the control panel of the
module where the error has occurred, and in the System Error Log. The System Error Log
automatically stores the 2000 most recent error messages in the order they occurred.
Aliquot Module Function Codes
The procedure to recover from an error condition can require that you enter one or more
function codes into one of the control panels on the Power Express or clearing an error
condition on a connected instrument.
The following table lists the Power Express module function codes by associated module,
with descriptions of the action triggered by each function code. You must place a module in
Manual mode to enter function codes.
Module Function Code Definitions
Table 3.1
Unit
Function
Code
Definition
Level
Detector
01
Lane through (Bypass)
82
Set Load Timeout Parameter to start the measurement of red cell layer.
91
Collect all sample tubes located downstream from BR01
Labeler
01
Lane through (Bypass)
91
Collect all sample tubes located downstream from BR02
Aliquotter
01
Lane through (Bypass)
31
Move tip nozzles toward the front of the module
32
Move tip nozzles toward the back of the module
Error Recovery Procedures
Error Recovery General Information
3-2
B31449AE
Содержание Power Express Aliquot
Страница 6: ...Warranty and Service vi B31449AE...
Страница 22: ...Introduction Scope of this Manual xxii B31449AE...
Страница 30: ...Power Express Aliquot Module Description Components 1 8 B31449AE...
Страница 110: ...Figure 4 3 Tube Supply Sensor Diagram 4 50 B31449AE Troubleshooting...
Страница 111: ...Figure 4 4 Label Supply Sensor Diagram B31449AE 4 51 Troubleshooting...
Страница 112: ...Aliquot Unit Figure 4 5 Overview Sensor Diagram 4 52 B31449AE Troubleshooting...
Страница 114: ...Figure 4 7 Tip Supply Sensor Diagram 4 54 B31449AE Troubleshooting...
Страница 117: ...Figure 4 10 Optional Track Sensor Diagram B31449AE 4 57 Troubleshooting...
Страница 118: ...Figure 4 11 Merge Lane Sensor Diagram 4 58 B31449AE Troubleshooting...
Страница 119: ...APPENDIX A Related Documents Related Documents General System Operation IFU PN B31448 B31449AE A 1...
Страница 120: ...Related Documents Related Documents A 2 B31449AE...
Страница 121: ......
Страница 122: ...www beckmancoulter com 2018 Beckman Coulter Inc All Rights Reserved...