Aliquotter Error Codes (Continued)
Table 4.4
Sensor
Problem
Solution
SL004A ON
SL004B ON
SL004 OFF
The cylinder used to extend
the drip tray under aliquot
arm A has an error.
ON error: The cylinder does
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
1. Inspect for and remove any obstructions.
2. Select
Restart
on the control panel to resume
operation.
3. Confirm that the air compressor is on and the auto switch
is working correctly.
4. Confirm that the compressed air supply is within the
normal pressure range.
SL011 ON
SL011 OFF
SL012 ON
SL012 OFF
SL013 ON
SL013 OFF
Stopper error. The stopper
is not working correctly.
ON error: The cylinder does
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
1. Confirm that a carrier is not jammed or stuck.
2. Confirm that the air compressor is on and the auto switch
is working correctly.
3. Confirm that the compressed air supply is within the
normal pressure range.
4. Select
Restart
on the control panel to resume
operation.
SL016A ON
SL016B ON
Error at the splash guard
cover at the aspiration
point.
1. Inspect for and remove any obstructions.
2. Confirm that the air compressor is on and the auto switch
is working correctly.
3. Confirm that the compressed air supply is within the
normal pressure range.
SL021 ON
SL021 OFF
SL022 ON
SL022 OFF
SL023 ON
SL023 OFF
Stopper error. The stopper
is not working correctly.
ON error: The cylinder does
not respond correctly to an
ON signal from a PLC.
OFF error: The cylinder
does not respond correctly
to an OFF signal from a PLC.
1. Confirm that a carrier is not jammed or stuck.
2. Confirm that the air compressor is on and the auto switch
is working correctly.
3. Confirm that the compressed air supply is within the
normal pressure range.
4. Select
Restart
on the control panel to resume
operation.
SL026A ON
SL026B ON
Error at the splash guard
cover at the front of the
module.
1. Inspect for and remove any obstructions.
2. Confirm that the air compressor is on and the auto switch
is working correctly.
3. Confirm that the compressed air supply is within the
normal pressure range.
Troubleshooting
Error Code Tables
4-30
B31449AE
Содержание Power Express Aliquot
Страница 6: ...Warranty and Service vi B31449AE...
Страница 22: ...Introduction Scope of this Manual xxii B31449AE...
Страница 30: ...Power Express Aliquot Module Description Components 1 8 B31449AE...
Страница 110: ...Figure 4 3 Tube Supply Sensor Diagram 4 50 B31449AE Troubleshooting...
Страница 111: ...Figure 4 4 Label Supply Sensor Diagram B31449AE 4 51 Troubleshooting...
Страница 112: ...Aliquot Unit Figure 4 5 Overview Sensor Diagram 4 52 B31449AE Troubleshooting...
Страница 114: ...Figure 4 7 Tip Supply Sensor Diagram 4 54 B31449AE Troubleshooting...
Страница 117: ...Figure 4 10 Optional Track Sensor Diagram B31449AE 4 57 Troubleshooting...
Страница 118: ...Figure 4 11 Merge Lane Sensor Diagram 4 58 B31449AE Troubleshooting...
Страница 119: ...APPENDIX A Related Documents Related Documents General System Operation IFU PN B31448 B31449AE A 1...
Страница 120: ...Related Documents Related Documents A 2 B31449AE...
Страница 121: ......
Страница 122: ...www beckmancoulter com 2018 Beckman Coulter Inc All Rights Reserved...