10
If there are no sample tubes at the Recapper, proceed to Step 14.
11
Check for caps in the Recapper lateral cap pusher:
a. Open the rear doors.
b. Look for caps in lateral cap pusher and plunger chute. If there are caps, proceed to
Step 12. If there are no caps, proceed to Step 13.
12
From the rear of the Recapper, remove caps from the lateral cap pusher:
a. If sample tubes are in the capping area, push them back on the track.
b. Remove all caps at the lateral cap pusher. If there is a cap in the plunger chute, use a
flat head screwdriver to push the cap down and out of the cylinder. Clear track of all
caps.
Important
If one or more caps stay in the lateral cap pusher, the system can generate an
error.
13
Select
Auto
and
OK
to place the module back into operation. Follow the
instructions in the Error Code section for any errors that occur at the Recapper.
14
In the Recapper:
If...
Then...
there is a sample tube
with a cap at SN003
or SN033 (FS002 ON,
FS004 ON, FS006 ON,
FS032 ON, FS034 ON,
or FS036 ON error)
Remove the tubes from the sample tube carrier at SN003 or SN033. When
the module is placed back into Auto mode, the empty sample tube carriers
continue on the track. Confirm that the cap is correctly placed and put the
tubes in the same sample carrier after startup.
there is a sample tube
without a cap at
SN003 or SN033.
(FS002 ON, FS004 ON,
FS006 ON, FS032 ON,
FS034 ON, or FS036
ON error)
Remove the sample tubes from the sample tube carrier at SN003 or SN033.
When the module is placed back into Auto mode, the empty sample tube
carriers continue on the track. Place cap securely on the tubes and put the
tubes in the same sample carriers after startup.
15
At the Stago Rack Builder modules:
a. Do not stop the connected analyzers. The analyzers continue to process sample
tubes.
b. If tubes are at BR02 or SN031, leave them in place.
c. If tubes are in the single arm gripper, Load, or Unload racks, remove the tubes.
d. Check the status of the samples on the analyzer for the following:
Error Recovery Procedures
Power Express Aliquot Error Recovery Procedures
3-10
B31449AE
Содержание Power Express Aliquot
Страница 6: ...Warranty and Service vi B31449AE...
Страница 22: ...Introduction Scope of this Manual xxii B31449AE...
Страница 30: ...Power Express Aliquot Module Description Components 1 8 B31449AE...
Страница 110: ...Figure 4 3 Tube Supply Sensor Diagram 4 50 B31449AE Troubleshooting...
Страница 111: ...Figure 4 4 Label Supply Sensor Diagram B31449AE 4 51 Troubleshooting...
Страница 112: ...Aliquot Unit Figure 4 5 Overview Sensor Diagram 4 52 B31449AE Troubleshooting...
Страница 114: ...Figure 4 7 Tip Supply Sensor Diagram 4 54 B31449AE Troubleshooting...
Страница 117: ...Figure 4 10 Optional Track Sensor Diagram B31449AE 4 57 Troubleshooting...
Страница 118: ...Figure 4 11 Merge Lane Sensor Diagram 4 58 B31449AE Troubleshooting...
Страница 119: ...APPENDIX A Related Documents Related Documents General System Operation IFU PN B31448 B31449AE A 1...
Страница 120: ...Related Documents Related Documents A 2 B31449AE...
Страница 121: ......
Страница 122: ...www beckmancoulter com 2018 Beckman Coulter Inc All Rights Reserved...