1
Do not stop the connected analyzers. The analyzers continue to process sample tubes.
2
Reset all
Stop
buttons that were pressed by pressing the buttons again. You cannot
restore power to the system until all
Stop
buttons have been reset.
3
Press the green
ON
button on the Inlet module to restore power to the system. Resolve
the conditions that prompted the emergency stop. The modules do not yet home or
process any sample tubes.
4
At the Inlet module, remove any tubes from the grippers and return them to the
appropriate Inlet rack:
a. If a stopper is jamming a sample tube carrier, push it forward to release the carrier.
b. If the transfer arm is extended, follow the procedure below to retract the arm to the
standby position.
— Select
Manual
and
OK
to place the module in Manual mode.
— Select
Maintenance
on the control panel.
— Select the number field to display the onscreen keyboard.
— Enter function code
06
and select
Enter
twice.
c. If there are sample tubes in the sample tube carriers in the tube loading area,
remove the tubes from the carriers.
d. Remove sample tubes in the sample tube carriers at the RF01 and RF02 positions.
e. Determine which rack or Bulk Inlet the samples removed in steps c and d were
loaded from. Reload the sample tubes to racks in the Inlet module or the Bulk Inlet
bin.
5
At the Error Lane, if a stopper is jamming the sample tube carrier, push it forward to
release the carrier.
6
At the Bulk Inlet module, remove sample tubes from the transfer mechanism if a
stopper is jamming the sample carrier.
a. If a tube is at FS011 or FS021, and it is not fully seated in the carrier, push the
sample tube down
b. If a tube is at FS131 or FS231, remove the tube and replace it in the loader bin.
Rotate DM131 or DM231 to remove the tube and return the motor to a horizontal
position.
c. If a stopper is jamming the sample tube carrier, push it forward to release the
carrier.
7
At the Outlet module, remove sample tube from the gripper.
a. If a sample tube is present in the gripper finger, manually remove the tube and
place it back into the empty sample tube carrier at BR01.
b. If the transfer arm is extended, follow the procedure below to retract the arm to the
standby position.
— Select
Maintenance
on the control panel.
— Select the number field to display the onscreen keyboard.
— Enter function code 06 and select
Enter
twice.
Error Recovery Procedures
Power Express Aliquot Error Recovery Procedures
3
B31449AE
3-7
Содержание Power Express Aliquot
Страница 6: ...Warranty and Service vi B31449AE...
Страница 22: ...Introduction Scope of this Manual xxii B31449AE...
Страница 30: ...Power Express Aliquot Module Description Components 1 8 B31449AE...
Страница 110: ...Figure 4 3 Tube Supply Sensor Diagram 4 50 B31449AE Troubleshooting...
Страница 111: ...Figure 4 4 Label Supply Sensor Diagram B31449AE 4 51 Troubleshooting...
Страница 112: ...Aliquot Unit Figure 4 5 Overview Sensor Diagram 4 52 B31449AE Troubleshooting...
Страница 114: ...Figure 4 7 Tip Supply Sensor Diagram 4 54 B31449AE Troubleshooting...
Страница 117: ...Figure 4 10 Optional Track Sensor Diagram B31449AE 4 57 Troubleshooting...
Страница 118: ...Figure 4 11 Merge Lane Sensor Diagram 4 58 B31449AE Troubleshooting...
Страница 119: ...APPENDIX A Related Documents Related Documents General System Operation IFU PN B31448 B31449AE A 1...
Страница 120: ...Related Documents Related Documents A 2 B31449AE...
Страница 121: ......
Страница 122: ...www beckmancoulter com 2018 Beckman Coulter Inc All Rights Reserved...