a. Check the area around the tip delivery lane.
b. If a stopper is jamming the sample tube carrier, push it forward to release the
carrier.
c. Check to see if there are jammed or dropped tips and remove if necessary.
d. Save samples that are being aliquotted. Check to see if the aliquotter tips have any
sample in them. Use the following procedure to save the sample:
1. Check to see if the Aliquot tips have any sample in them.
2. Find the sample ID of the primary sample, at the Line Control, select
System
Setup
>
Aliquotting Sample ID
.
The system displays sample IDs. Arm #1 is associated with the right Aliquot
Nozzle if viewed from the front of the Aliquot component. Arm #2 is the left
Aliquot Nozzle.
3. Move the aliquot tip nozzles toward the back of the module with Aliquot mode
32 as follows:
1. Select
Maintenance
on the front control panel of the Aliquot component.
2. Select the number field to display the onscreen keyboard.
3. Enter code 32 and select
Enter
twice on the control panel.
4. If TIP ERR 12, TIP ERR 13, or TIP ERR 23 occurs, select
Clear
and
OK
on the control panel. If the
Clear
option is not displayed, dispose of the
aliquot tip down the disposal chute.
4. Unlock the rear doors of the Aliquotter and open the rear doors.
5. Collect the sample by placing a 2ml sample cup below aliquot tip #1 and then
pulling the aliquot tip downward. Repeat for nozzle #2.
6. After collecting the sample, re-attach the aliquot tip to the nozzle.
7. Close the Aliquotter rear safety shield doors.
e. Use the following procedure to remove any tubes on the conveyor lane.
1. Remove any tubes on the Aliquotter and Labeler conveyor lane.
a. On the Aliquotter control panel, select
Maintenance
.
b. Select
Bypass
for the tubes to bypass the Aliquot module.
c. On the Labeler control panel, select
Maintenance
to display the
Maintenance settings.
d. Select the number field to display the onscreen keyboard.
e. Enter Function code 91 and select
Enter
twice to hold tubes at bar code
reader BR02 and to release tubes after BR02.
f. Remove all tubes in the Aliquotter and the Labeler. Do not remove any
sample tubes at BR02 or before the tube is in the Labeler.
2. After all the tubes and sample carriers have left the Aliquotter and Labeler, exit
the function codes and return the module to Auto mode:
a. On the Aliquotter control panel, select
Enter
and
Exit
.
Error Recovery Procedures
Power Express Aliquot Error Recovery Procedures
3-14
B31449AE
Содержание Power Express Aliquot
Страница 6: ...Warranty and Service vi B31449AE...
Страница 22: ...Introduction Scope of this Manual xxii B31449AE...
Страница 30: ...Power Express Aliquot Module Description Components 1 8 B31449AE...
Страница 110: ...Figure 4 3 Tube Supply Sensor Diagram 4 50 B31449AE Troubleshooting...
Страница 111: ...Figure 4 4 Label Supply Sensor Diagram B31449AE 4 51 Troubleshooting...
Страница 112: ...Aliquot Unit Figure 4 5 Overview Sensor Diagram 4 52 B31449AE Troubleshooting...
Страница 114: ...Figure 4 7 Tip Supply Sensor Diagram 4 54 B31449AE Troubleshooting...
Страница 117: ...Figure 4 10 Optional Track Sensor Diagram B31449AE 4 57 Troubleshooting...
Страница 118: ...Figure 4 11 Merge Lane Sensor Diagram 4 58 B31449AE Troubleshooting...
Страница 119: ...APPENDIX A Related Documents Related Documents General System Operation IFU PN B31448 B31449AE A 1...
Страница 120: ...Related Documents Related Documents A 2 B31449AE...
Страница 121: ......
Страница 122: ...www beckmancoulter com 2018 Beckman Coulter Inc All Rights Reserved...