9
Close the Aliquotter rear safety shield doors.
10
Perform the
Aliquotter and Labeler Lane Clear
procedure.
Aliquotter and Labeler Lane Clear
Use this procedure to clear any sample tubes and sample carriers after the aliquot process
has been canceled. This includes:
• After a Stop Button has been pressed.
• After an ID error is detected in the Aliquot process (primary and secondary bar code
ID’s do not match).
• After a Drip Tray error in which the drip trays are blocked or otherwise inaccessible.
1
Remove any tubes on the Aliquotter and Labeler conveyor lane.
a. On the Aliquotter control panel, select
Maintenance
.
b. Select
Bypass
for the tubes to bypass the Aliquot module.
c. On the Labeler control panel, select
Maintenance
to display the Maintenance
settings.
d. Select the number field to display the onscreen keyboard.
e. Enter Function code 91 and select
Enter
twice to hold tubes at bar code reader
BR02 and to release tubes after BR02.
f. Remove all tubes in the Aliquotter and the Labeler. Do not remove any sample tubes
at BR02 or before the tube is in the Labeler.
2
After all the tubes and sample carriers have left the Aliquotter and Labeler, exit the
function codes and return the module to Auto mode:
a. On the Aliquotter control panel, select
Enter
and
Exit
.
b. On the Labeler control panel, select
Enter
and
Exit
.
c. Select
Auto
and
OK
on both the Aliquotter and Labeler to resume
processing.
Aliquot Sample Recovery
If necessary, use this procedure to recover any sample left in the aliquot tips after the
aliquot process has been cancelled. This includes:
• After a Stop Button has been pressed.
• After an ID error is detected in the Aliquot process (primary and secondary bar code
ID’s do not match).
• After a Drip Tray error in which the drip trays are blocked or otherwise inaccessible.
1
Check to see if the Aliquotter tips have any sample in them.
Error Recovery Procedures
Aliquot Module Error Recovery Procedures
3
B31449AE
3-19
Содержание Power Express Aliquot
Страница 6: ...Warranty and Service vi B31449AE...
Страница 22: ...Introduction Scope of this Manual xxii B31449AE...
Страница 30: ...Power Express Aliquot Module Description Components 1 8 B31449AE...
Страница 110: ...Figure 4 3 Tube Supply Sensor Diagram 4 50 B31449AE Troubleshooting...
Страница 111: ...Figure 4 4 Label Supply Sensor Diagram B31449AE 4 51 Troubleshooting...
Страница 112: ...Aliquot Unit Figure 4 5 Overview Sensor Diagram 4 52 B31449AE Troubleshooting...
Страница 114: ...Figure 4 7 Tip Supply Sensor Diagram 4 54 B31449AE Troubleshooting...
Страница 117: ...Figure 4 10 Optional Track Sensor Diagram B31449AE 4 57 Troubleshooting...
Страница 118: ...Figure 4 11 Merge Lane Sensor Diagram 4 58 B31449AE Troubleshooting...
Страница 119: ...APPENDIX A Related Documents Related Documents General System Operation IFU PN B31448 B31449AE A 1...
Страница 120: ...Related Documents Related Documents A 2 B31449AE...
Страница 121: ......
Страница 122: ...www beckmancoulter com 2018 Beckman Coulter Inc All Rights Reserved...