Commissioning/Configuration
EP3356-0022
43
Version: 1.4
9. Load the scales with a reference load. This should be at least 20 % of the rated load. The larger the
reference load, the better the sensor values can be calculated. In object
("Reference
load"), enter the load in the same unit as the rated load (
value indicates a constant value for at least 10 seconds, execute the command "0x0102" (258dec) on
CoE object
. By means of this command the EP3356 determines the output value for
the nominal weight ("Rated output") Check: CoE objects
and
contain "0" after execution.
10. Reset: execute the command "0x0000" (0dec) on CoE object
11. Set the filter to a lower stage.
Calibration according to the sensor calibration protocol (theoretical calibration)
The sensor characteristic values according to the manufacturer's certificate are communicated here directly
to the EP3356, so that it can calculate the load.
1. Perform a CoE reset with object 0x1011:01, see
Restoring the delivery state [
) (for declaration of weight in g)
3. Set gravity of earth (
) if necessary (default: 9.806650)
6. Specify the nominal load of the sensor in
("Nominal load")
7. Adopt the "Rated output" (mV/V value
) from the calibration protocol
) from the calibration protocol
Calibration
The calibration is of great importance for the accuracy of the system. In order to increase this, the
filter should be set as strong as possible over the entire calibration phase. It may take several sec-
onds before a static value is obtained.
Local storage
The values modified during the theoretical and practical calibration are stored in a local EEPROM.
This can be written to up to 1 million time. In order to prolong the life of the EEPROM, therefore, the
commands should not be executed cyclically.