Alaris™ GP (Guardrails™) Volumetric Pump
Getting Started
BDDF00535
Issue 1
17/56
Step
Notes
6 . Close roller clamp .
7 . Open door and load infusion set as follows:
• Ensure infusion set tubing above the Pump is as
straight as possible with no kinks
Top set retainer
(Blue)
Adaptor on
infusion set
(Blue)
Pumping
mechanism
Safety Clamp
retainer (Orange)
Air-in-line sensor
Tubing guide
Alaris Safety
Clamp (Orange)
i) Fit top adaptor of infusion set into top set retainer .
Top Set Retainer
ii) Insert Safety Clamp into retainer .
• Avoid any stretching of the silicone segment when
loading, priming and re-loading the infusion set
Safety Clamp
w
Pushing on the Safety Clamp Slider may lead
to uncontrolled flow to the patient . Therefore,
always close the roller clamp before pushing
on the safety clamp slider .