Alaris™ GP (Guardrails™) Volumetric Pump
Service Configuration Mode
BDDF00535
Issue 1
28/56
Service Configuration Mode
This section comprises of a list of options which can be configured . Some can be entered via the Pump
SERVICE
menu (available in
Technician Mode) and others through the Alaris Editor Software .
w
Access codes should only be entered by Qualified Service Personnel .
Alarm Presets
Pumps with version 2 .3 .4 software have 2 alarm tones to choose from during configuration:
•
ORIGINAL ALARMS
: Low, medium and high priority alarm tones that sound like the auditory alarms and warnings from software
versions prior to 2 .3 .4
•
3RD EDITION ALARMS
: Low, medium and high priority alarm tones in accordance with IEC 60601-1-8: 2012 and IEC 60601-2-
24:2012
Enter the access code on the Pump for Alarm Presets, see the
Technical Service Manual
or
Information Notice
for details .
1 . Use the
f
keys to select alternative alarm tones .
2 . When the desired alarm tone has been selected press
OK
softkey .
3 . When all modifications have been carried out press
QUIT
softkey .
w
All pumps in a single care area should be configured with the same alarm tones to avoid User confusion .
The Hospital/Facility is responsible for selecting and configuring the desired alarm scheme .
Alaris Gateway Workstation (Workstation) with software versions 1 .1 .3, 1 .1 .3 MR, 1 .1 .5, 1 .2, 1 .3 .0, 1 .6 .0 or 1 .5 do not
support the new Pump low priority visual alarms scheme defined in IEC 60601-1-8: 2012 . For Pumps with version 2 .3 .4
software or higher docked into these Workstations there will be a mismatch of alarm priority displayed . As a result,
Near End Of Infusion, AC Power Fail, Add Drug Not Complete, and Attention alarms will display as a medium visual
priority alarms on the Workstation beacon and a low priority alarm on the Pump . Additionally, for certain information
signals, e .g . those associated with Add Drug Not Complete and Titration Not Confirmed, the Workstation beacon will
illuminate while the beacon on the pump will not . In the event of an alarm priority mismatch, the User should refer to
the alarm on the Pump for the correct priority .
Configured Options
Enter the access code on the Pump for
SERVICE
mode, then select
CONFIGURATION
, see the Technical Service Manual for details .
Use Alaris Editor to configure the Pump configuration, drug library and units enabled for each data set .
Date & Time
1 . Select
DATE & TIME
from the
CONFIGURATION
menu using the
f
keys and press the
OK
softkey .
2 . Press the
OK
softkey to confirm .
3 . Use the
f
keys to adjust the date displayed, pressing the
NEXT
softkey to access the next field .
4 . When the correct date and time are displayed press the
OK
softkey to return to the
CONFIGURATION
menu .
5 . Press the
QUIT
softkey to return to the
SERVICE
menu and press
a
to exit and power down .
Pump Reference
This option is used to add reference text to be shown on the Pump start up display .
1 . Select
PUMP REFERENCE
from the
CONFIGURATION
menu using the
f
keys and press the
OK
softkey .
2 . Use the
f
keys to enter the text and
NEXT
to move to the next character .
3 . When the desired text has been selected press
OK
softkey to return to the
CONFIGURATION
menu .
4 . Press
QUIT
to exit back to the main
SERVICE
menu and press
a
to exit and power down .
Language
This option is used to set the language of messages shown on the Pump display .
1 . Select
LANGUAGE
from the
CONFIGURATION
menu using the
f
keys and press the
OK
softkey .
2 . Use the
f
keys to select the language .
3 . When the desired language has been selected press
OK
softkey to return to the
CONFIGURATION
menu .
4 . Press
QUIT
to exit back to the main
SERVICE
menu and press
a
to exit and power down .