Alaris™ GP (Guardrails™) Volumetric Pump
Basic Features
BDDF00535
Issue 1
25/56
Primary Setup
If a secondary infusion has already been setup (See 'Secondary (Piggyback) Infusions' section), then access to the primary infusion setup
is as follows:
1 . Press
h
to put the Pump
ON HOLD
.
2 . Press the
d
button to access the options menu .
3 . Select
PRIMARY Setup
and press the
OK
softkey to confirm . Make changes to the primary setup as necessary .
Secondary Setup
To setup a secondary infusion refer to 'Secondary (Piggyback) Infusions' section
Setting VTBI over Time
This option allows a specific VTBI and delivery time (Maximum of 24 hours) to be set . The rate necessary to deliver the required volume
within the specified time is calculated and displayed .
1 . Stop the infusion . Press
d
button to access the options menu .
2 . Select the
SET VTBI OVER TIME
option using the
f
keys and press the
OK
softkey .
3 . Adjust the volume to be infused using the
f
keys (Or select
BAGS
softkey to set the VTBI) . When the desired volume has
been reached press the
OK
softkey .
4 . Enter the time over which the volume is to be infused using the
f
keys . The infusion rate will automatically be calculated .
5 . Press
OK
softkey to enter the value or
BACK
to return to the VTBI .
Adjust Alarm Volume
This option allows adjustment of the volume if enabled .
1 . Press the
d
button to access the options menu .
2 . Select
ADJUST ALARM VOLUME
using the
f
keys and press
OK
softkey to confirm .
3 . Select
HIGH
,
MEDIUM
or
LOW
using the
f
keys .
4 . Press
OK
softkey to confirm or
QUIT
to exit screen .
Event Log
This option allows the event log to be reviewed . The event log holds up to 99,960 events . When the log is full, the oldest events will be
overwritten by the newest occurring events .
1 . Press the
d
button to access the options menu .
2 . Select
EVENT LOG
using the
f
keys and press
OK
softkey to confirm .
3 . View the events using the
f
keys .
4 . Select
BACK
to return to the previous screen, if required .
Pump Details
To review Pump information:
1 . Press the
d
button to access the options menu .
2 . Select
PUMP DETAILS
using the
f
keys and press
OK
softkey to confirm .
3 . Review the information and then press the
QUIT
softkey .
Profile Filter (Only available on the Alaris GP Guardrails Volumetric Pump)
This option allows selected profiles to be filtered from view . It can be enabled/disabled via the Alaris Editor .
1 . Press the
d
button to access the options menu .
2 . Select
PROFILE FILTER
using the
f
keys, press the
OK
softkey to confirm .
3 . Select profile that requires filtering using the
f
keys .
4 . Press
MODIFY
to disable profile and press
OK
to confirm or
QUIT
to return to main display .
Note:
If only one profile is available and all others are disabled in this option, the
CONFIRM PROFILE
screen will not display upon
start up .
Standby
This option allows the Pump to be placed on standby mode . It can be enabled/disabled via the Alaris Editor .
1 . Press the
d
button to access the options menu .
2 . Select
STANDBY
using the
f
keys, press the
OK
softkey to confirm .
3 . Select
CANCEL
to return to main display .