Alaris™ GP (Guardrails™) Volumetric Pump
Main Display Features
BDDF00535
Issue 1
9/56
Main Display Features
Main Display - If VTBI has not been set (flow sensor must be used)
Infusion Status /
Drug Name / Profile Name /
Primary or Secondary (Only if
secondary is enabled in the
data set)
Infusion Rate
Volume Infused
Softkeys
Softkey Identifiers
Clear Volume Infused Set VTBI Option
ON HOLD
RATE
VOLUME
VTBI
ml/h
VOLUME
ml
25.0
50
.0
a)
ON HOLD
RATE
SET RATE WITH
ON HOLD
RATE
RATE TOO LOW
ON HOLD
RATE
RATE TOO HIGH
If the rate has not been set and is
showing at 0 .0ml/h, then message
a)
will be displayed .
b)
ON HOLD
RATE
SET RATE WITH
ON HOLD
RATE
RATE TOO LOW
ON HOLD
RATE
RATE TOO HIGH
If programmed rate is between 0 .0ml/h
and 0 .1ml/h exclusive in drug protocol,
message
b
) will be shown .
Main Display - If VTBI is set
Infusion Status / Drug Name
/ Profile Name / Primary or
Secondary (Only if secondary
is enabled in the data set)
Infusion Rate
Dose Rate
Volume to be Infused
Volume Infused
Time remaining
Softkeys
Softkey Identifiers
Clear Volume Infused
Set VTBI Option
ADRENALINE
RATE
VOLUME
VTBI
25
.0
ml/h
µg/kg/24h
16.
7
1h 48m 00s
VTBI
VOLUME
ml
ml
45
.0
50
.0
c)
ON HOLD
RATE
SET RATE WITH
ON HOLD
RATE
RATE TOO LOW
ON HOLD
RATE
RATE TOO HIGH
If programmed rate is greater than the
Infusion Rate Max
in drug protocol,
message
c)
will be shown .
Screen Icons
Symbol
Description
Time remaining display icon - Indicates time remaining before VTBI will be completed . If the time is greater than
24 hours then 24+ will be displayed .
N
Battery icon - Indicates battery charge level to highlight when the battery will require recharging .
Pressure Information icon - Shows the pressure from level 0 being the first bar to level 8 . Alarm limits: level 0 - 8 .
?
Indicates that the value entered is outside of the Guardrails soft limits . The warning can be overridden (Indicates
Guardrails safety protocol is in use) .
!
Indicates that the value entered is outside of the hard limits . The warning can NOT be overridden . This symbol is
also used to prompt the user to set the rate .
Indicates that the Pump is running at a rate below (pointing down) a Guardrails Soft Limit .
Indicates that the Pump is running at a rate above (pointing up) a Guardrails Soft Limit .