Alaris™ GP (Guardrails™) Volumetric Pump
Flow Sensor Operation (Optional)
BDDF00535
Issue 1
40/56
Flow Sensor Operation (Optional)
w
• The flow sensor automatically monitors the infusion flow rate through the drip chamber . The flow sensor will
cause the Pump to alarm if a significant deviation from the infusion rate occurs . The flow sensor will also be able
to detect empty containers . For this reason we recommend use of a flow sensor wherever possible excluding
secondary infusions .
• When infusing critical drugs, it is recommended to use a flow sensor, in addition to entering a VTBI .
Alaris®
Guardrails
®
GP
Alaris
®
Guardrails
®
GP
plus
IVAC™ Model 180 Flow Sensor
Handles
Drip
chamber
Flanges
Pull back handles
Flow sensor
interface
1 . Plug the flow sensor into the flow sensor interface located on the top rear part of the Pump .
2 . Attach the IVAC™ Model 180 Flow Sensor to the drip chamber of the infusion set, by pulling back the handles . Refer to the illustration
above .
3 . Proceed with load, priming, and set-up instructions as described in section 'Getting Started' .
Note:
Ensure drip chamber is half full and upright .
w
Always attach the flow sensor before you start an infusion .
Avoid using the flow sensor in direct sunlight .
Always ensure lens is clean .
Flow
Sensor
Interface
Cover
w
Always replace the flow sensor interface cover when the flow sensor is disconnected .