Alaris™ GP (Guardrails™) Volumetric Pump
Specifications
BDDF00535
Issue 1
48/56
Specifications
Electrical Protection
Class I, Type CF (Defibrillation-proof)
Electrical/Mechanical Safety
Complies with IEC/EN60601-1 and IEC/EN60601-2-24 .
Electro Magnetic Compatibility (EMC)
Complies with IEC/EN60601-1-2 and IEC/EN60601-2-24 .
Electrical Safety
Typical earth leakage current 78µA .
Typical Enclosure Leakage Current (Normal Condition) = 0µA
Typical Protective Earth Resistance = 32mOhms
The above measurements are for guidance only, IEC/EN60601-1 limits are defined below:
Earth Leakage Current (Normal Condition) <= 500µA
Enclosure Leakage Current (Normal Condition) <= 100µA
Protective Earth Resistance <= 200mOhms
Classification
Continuous mode of operation, Portable Equipment
AC Power Supply
100 - 230 VAC, 50 - 60Hz, 60VA (Maximum) .
Fuse Type
2 X T 1 .25 A, slow blowing .
Dimensions
148mm (w) x 225mm (h) x 148mm (d) . Weight: approx . 2 .5kg (excluding power cable) .
Protection against fluid ingress
IP33 - Protected against direct sprays up to 60° from the vertical .
Environmental Specifications
Condition
Operating
Transport & Storage
Temperature
+0°C - +40°C
-20°C - +50°C
Humidity
20% - 90%*
15% - 95%*
Atmospheric Pressure
700hPa - 1060 hPa
500hPa - 1060hPa
*Non condensing .
Battery Specifications
Rechargeable NiMH (Nickel Metal Hydride) . Automatically charges when the Pump is connected to AC power .
• Battery Charging - 2 .5 hours to 95% .
Battery Life
Infusion Rate
Mean Time To Battery Empty from fully charged
25ml/h
5 .8
hours
125ml/h
5 .65 hours
1200ml/h
3 .4 hours