Ipump Pain Management System Operator’s Manual
5-35
CD07-19-E4-766
Changing the Prescription During Infusion
The prescription and units cannot be changed unless the pump is turned off and then reprogrammed. If it is
necessary to retain a patient’s history information, you can review and record the history data on the patient’s
chart or print a hard copy before proceeding with this procedure. See “Reviewing the Therapy History,” 5-37.
To program a supplemental (clinician) bolus, see “Programming a Supplemental (Clinician) Bolus,” 5-28.
Action
Message Displayed
1.
Press
STOP
twice to interrupt the infusion or bolus.
KEY+CODE or KEY ONLY:
PUMP READY
PRESS START
CODE ONLY:
PUMP READY
START OR CLEAR