Ipump Pain Management System Operator’s Manual
9-1
CD07-19-E4-766
Chapter 9. Technical Specifications
Component
Description
AC Power Requirements (when used with optional
AC adapter)
With 2L3210 adapter: 100 to 120 VAC 60 Hz
With 2L3205K adapter: 220 to 240 VAC 50 Hz, 700mA
DC Power Requirements
9V alkaline battery
Typical operating time when operating at an intermediate rate of 1 mL/hr is approximately
140 hours. Typical operating time when operating at a rate of 10 mL/hr is approximately
76 hours.
Leakage Current
Less than 0.3 m
Α
earth leakage (tested per UL 2601-1)
AC Adapter Cord (100-120V) — US
Approximately 1.8 m (5.9 feet) long
AC Adapter Cord (220-240V) — UK/Europe
Approximately 2.0 m (6.5 feet) long
Range of Programmable Flow Rates
0.1 to 90.0 mL/hr in 0.1 mL/hr increments
Maximum Infusion Under Single Fault Conditions 0.50 mL
Flow Rates
Basal Rate: 0.1 to 19.9 mL/hr
Continuous Rate: 0.1 to 90.0 mL/hr
PCA Dose, Bolus, and Priming Rate: 90.0 mL/hr