Ipump Pain Management System Operator’s Manual
3-16
CD07-19-E4-766
Note:
Whenever the pump is stopped, it automatically closes off the tubing to reduce the risk of a free flow condition. When the
tubing is removed from the pump, the anti-siphon valve on the set reduces the risk of a free flow condition when used in
accordance with the set’s instructions.
Filling and Loading the Fluid Bag
1.
Open the package and carefully remove the bag.
2.
For Baxter empty luer-lock 100 mL or 250 mL bags:
Note:
Do not confuse the nonvented cap (which is supplied in its own package) with the cap attached to the outlet tubing.
The cap attached to the outlet tubing is not airtight and will not prevent fluid leakage.
Q
Fill a sterile syringe with the solution to be placed in the bag.
Q
Remove and discard the cap from the outlet tubing of the bag.
Q
Connect the syringe to the female luer-lock fitting on the bag’s outlet tubing. (Do not use a needle.)
CAUTION
All luer-lock connections must be properly tightened. Over tightening may crack the luer and
cause leakage.
CAUTION
Follow any directions provided by the manufacturer of the fluid bag being used.
Oct02_0719E4766.book Page 16 Friday, August 3, 2007 1:46 PM