Ipump Pain Management System Operator’s Manual
9-2
CD07-19-E4-766
Flow Rate Accuracy
± 8% at 22°C to 23°C (71.6°F to 73.4°F) nominal; ±10% at the temperature extremes of
10°C (50°F) and 40°C (104°F). See "Environmental Operating Limits," 9-4.
Time to Detect Downstream Occlusions
Occlusion alarm delay at 0.1 mL/hr is 2.5 hours (maximum)
Occlusion alarm delay at 5 mL/hr is 5.5 minutes (maximum)
Bolus Volume Released after Downstream
Occlusions are Corrected
Bolus volume after occlusion is 0.5 mL (maximum)
Air-in-Line
Air-in-line alarm configured for HIGH sensitivity senses approximately 0.1 mL of air
Air-in-line alarm configured for LOW sensitivity senses approximately 0.5 mL of air
Maximum Infusion Pressure Generated
The maximum infusion pressure generated prior to alarm activation is 2109 mmHg (41 psi)
Downstream Occlusion Alarm Pressure
The downstream occlusion alarm pressure threshold is 1138 ± 517 mmHg (22 ± 10 psi)
Operational Features
PCA Dose Volume Selections: 0.0 to 9.9 mL (in 0.1 mL increments)
Bolus Volume Selections: 0.0 to 49.9 mL (in 0.1 mL increments)
Bag Volume Selections: 1 to 1999 mL
Delay Time Selections: 1 to 240 minutes
History/Prescription Recall
Component
Description