Ipump Pain Management System Operator’s Manual
5-31
CD07-19-E4-766
Starting, Stopping, Restarting the Infusion
Starting the Infusion
Action
Message Displayed
1.
Press
START
. The automatic upstream occlusion test is performed (see “Upstream
During the test, the
Alert
LED flashes red.
Upon successful completion of the test, the
infusion
starts, and the
Infusing
LED flashes green.
NOTE:
The pump will not begin infusing until it has successfully passed the upstream
occlusion test. If an
UPSTREAM OCCLUSION
or
SET NOT PRIMED
alarm occurs, the
test must be repeated until it passes successfully. See Table 6-2, "Alarm Messages and
Responses," on pages 6-18 and 6-15 for more information.
NOTE:
If the pump is configured for an automatic start after bolus, the infusion will begin
immediately after the delivery of the bolus if one has been programmed.
TESTING
UP OCCLUSION
Then:
PCA
BASAL+PCA
CONTINUOUS
00.0 mg/hr
Oct02_0719E4766.book Page 31 Friday, August 3, 2007 1:46 PM