Ipump Pain Management System Operator’s Manual
6-6
CD07-19-E4-766
BOLUS DONE
Situation: A bolus has completed, the pump is configured for manual start after bolus,
and neither the
START
or the
Enter
key has been pressed for 1 minute.
Action: Press
Enter
or
START
to start the infusion.
CODE
INCORRECT
Situation: An invalid security code has been entered.
Action: Enter the correct code.
FLUID VOLUME
IS LOW
Situation: The calculated fluid volume remaining in the bag equals or is less than the
amount required for configured time parameters.
Action: Prepare to replace the fluid bag if necessary.
LOW BATTERY
Situation: The battery power is low. A repeating alert tone is sounded. Remaining
battery life is approximately 6 hours or less when running the pump at 1 mL/hr.
Action:
1.
Press
Clear/Silence
during the alert to silence the audio for 60 minutes,
regardless of the Alert Silencing Time configuration setting.
2.
Replace the battery as soon as possible as specified in “Installing and Changing
the Battery,” 3-2.
Table 6-1
Alert Messages and Responses — continued
Alert Message
Situation/Action
Oct02_0719E4766.book Page 6 Friday, August 3, 2007 1:46 PM