Ipump Pain Management System Operator’s Manual
10-8
CD07-19-E4-766
Table 10-3 Guidance and Manufacturer's Declaration –
Electromagnetic Immunity- for Life-Supporting Equipment and Systems
The pump is intended for use in the electromagnetic environment specified below. The customer or the user
of the pump should assure that it is used in such an environment.
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should
be used no closer to any part of the pump, including cables,
than the recommended separation distance calculated from
the equation applicable to the frequency of the transmitter.
Recommended separation distance:
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz TO 80 MHz
outside ISM bands
a
3 Vrms
d=1.2
√
P
10 Vrms
150 kHz to 80 MHz
in ISM bands
b
3 Vrms (3)
d=4
√
P