Ipump Pain Management System Operator’s Manual
6-12
CD07-19-E4-766
CODE
INCORRECT
Situation: Three invalid security codes have been entered.
Action:
1.
If the pump is configured as KEY+CODE, unlock the cover, lock it again, and
enter the security code.
2.
If it is configured as CODE ONLY, turn off the pump and turn it on again.
COMMUNICATION
TIMEOUT
Situation: The configuration mode transfer was incomplete or was interrupted.
Action: Press
Clear/Silence
to cancel the alarm and return to the
START TO PROCEED CLEAR TO CANCEL
display.
COVER IS
UNLOCKED
Situation: The cover is unlocked while the pump is active and the pump is configured
for KEY+CODE or KEY ONLY.
Action: Lock the cover, and press the
START
key to resume the infusion.
Table 6-2
Alarm Messages and Responses — continued
Alarm Message
Situation/Action