Ipump Pain Management System Operator’s Manual
3-14
CD07-19-E4-766
Loading the Tubing Set
1.
Unlock and open the cover as described in “Unlocking and
Opening the Cover,” 3-10.
2.
Release the tubing door by sliding the silver latch toward the
side of the pump as indicated by the arrow.
3.
Open the tubing door by pulling it down.
4.
Load the tubing set into the groove in the tubing door.
Ensure the sensor tab (see
INSET
) is properly positioned.
! WARNING !
Only use sets manufactured by Baxter as
specified in “Accessories, Disposables, and
Recommended Sets,” 8-1.
! WARNING !
Avoid getting any liquids on the tubing set, inside
the tubing door, or in the tubing channel. Air
sensor functioning could be compromised or
permanent damage may result.
CAUTION
Ensure that the set is oriented as shown in Figure
3-7 and Figure 3-8. The pump will not work if the
sensor tab is missing or misaligned.
2
4
3
INSET
Figure 3-7 Load Tubing Set