Ipump Pain Management System Operator’s Manual
3-18
CD07-19-E4-766
4.
Connect the distal end (longer segment) of the pump tubing set to the patient’s access site, making certain
that the luer-lock connection is properly tightened.
Note:
To prevent upstream occlusions, open any optional clamp on the bag upstream of the pump before starting the infusion.
5.
After connecting the bag to the pump tubing set, place the tubing in the groove in the tubing door (see
“Loading the Tubing Set,” 3-14) then close the tubing door.
6.
Verify that the tubing door is properly closed and latched.
7.
For Spiked Bag & Set Loading:
Take the spiked bag and fold it about 1/3 of the way from the spike port.
Place the bag in the cover with the set pulling across the hinge of the pump and the spike pointing towards
the lock or upper corner above the lock (see Figure 3-9).
For Luer Lock Bag and Set Loading:
Place the Luer Lock bag in the cover with the Luer Lock connector
closest to the pump (see Figure 3-9).
8.
Pull the pump body over the bag cover making sure the set is not pinched in the cover in any way.
! WARNING !
The tubing set MUST NOT be connected to the patient while priming.
! WARNING !
Failure to latch the tubing door properly may result in a no flow condition.
Oct02_0719E4766.book Page 18 Friday, August 3, 2007 1:46 PM