Ipump Pain Management System Operator’s Manual
3-22
CD07-19-E4-766
4.
Mount the clamp to a stable pole or vertical rail that is 1.3 cm to 3.2 cm (0.5" to 1.25") in diameter and
tighten it. If the clamp is detached from the pump, make certain the arrow on the clamp is pointing up.
Note:
The pump must be in the clamp before locking the clamp. Fail-
ure to do so makes it possible for the pump to be removed with-
out using a key.
5.
Lock the clamp by inserting the Pole Clamp key (not
the Lock Box key), pushing it in, and rotating it
clockwise to the locked position.
Although the clamp may be tightened when it is locked,
it cannot be loosened enough to remove it.
Figure 3-11 Locking the Clamp
Oct02_0719E4766.book Page 22 Friday, August 3, 2007 1:46 PM