Ipump Pain Management System Operator’s Manual
5-28
CD07-19-E4-766
Programming a Supplemental (Clinician) Bolus
If a bolus is interrupted, it cannot be restarted automatically. To administer additional bolus volumes, you
must reprogram the bolus.
Action
Message Displayed
1.
Press
STOP
twice to interrupt the infusion or bolus.
KEY+CODE or KEY ONLY:
PUMP READY
PRESS START
CODE ONLY:
PUMP READY
START OR CLEAR
2.
If the pump is configured as KEY+CODE or KEY ONLY, unlock and lock the
fluid bag cover.
If the pump is configured as CODE ONLY, press
Clear/Silence.
If the pump is configured as KEY+CODE or CODE ONLY, enter the correct
security code at the
ENTER CODE OR START
prompt, and press
Enter
.
PUMP READY
PRESS START
PUMP READY
START OR CLEAR
000 ENTER CODE
↑
OR START