BANTE900P PORTABLE MULTIPARAMETER WATER QUALITY METER
41
Specifications
MODEL
BANTE900P
pH
Range
-2.000~20.000pH
Accuracy
0.002pH
Resolution
0.1, 0.01, 0.001pH
Calibration Points
1 to 5 points, USA, NIST, DIN Standards
Temperature Compensation
0~100°C, 32~212°F, Manual or Automatic
mV
Range
-1999.9~1999.9mV
Accuracy
±0.2mV
Resolution
0.1, 1mV
Calibration Points
1 point (Only for Relative mV mode)
Calibration Range
±200mV
Ion
Range
0.001~19999ppm, mg/L, mol/L (Depending on range of Ion Selective Electrode)
Accuracy
±0.5% F.S (Monovalent), ±1% F.S (Divalent)
Resolution
0.001, 0.01, 0.1, 1
Calibration Points
2 to 5 points
Calibration Solutions
0.001, 0.01, 0.1, 1, 10, 100, 1000, 10000ppm, mol/L, mg/L
Conductivity
Range
0~20.00, 200.0, 2000µS/cm, 20.00, 200.0mS/cm
Accuracy
±0.5% F.S
Calibration Points
1 to 5 points
Calibration Solutions
10µS/cm, 84µS/cm,1413µS/cm,12.88mS/cm,111.8mS/cm
Temperature Compensation
0~100°C, 32~212°F, Manual or Automatic
Temperature Coefficient
0.0~10.0%/°C
Compensation Modes
Linear or Pure Water
Cell Constant
K=0.1, 1, 10
Normalization Temperature
20 or 25°C
TDS
Range
0~10ppt (Max. 20ppt, depending on factor setting)
Accuracy
±1% F.S
TDS Factor
0.1~1.0 (Default 0.5)