BANTE900P PORTABLE MULTIPARAMETER WATER QUALITY METER
23
1.4
Press ENTER key to confirm, “Calibration” indicator begin
s flashing.
1.5 Wait for the measured value to stabilize, the display shows
“END”. The
meter returns to measurement mode automatically.
Calibration is completed.
MULTI-POINT CALIBRATION:
2.1 Make sure that you have selected 2 to 5 points calibration in setup menu.
2.2 Repeat steps 1.2 to 1.5 above, when the first calibration point is completed, the display shows
“CAL2”. The
meter prompts you
to continue with second point calibration.
2.3 Rinse the conductivity electrode with distilled water.
Enter