BANTE900P PORTABLE MULTIPARAMETER WATER QUALITY METER
27
DO Calibration in % Saturation Mode
Bante900P portable meter can be calibrated quickly in air. In the percentage saturation mode, the meter is able to perform either 1 or 2
points calibration. For single point calibration, we recommend that you perform 100% saturation calibration in the air. If you select 2
points calibration, you need to use zero oxygen solution.
100% SATURATION CALIBRATION:
1.1 Press MODE key until the meter shows measurement unit
“%”.
1.2 Make sure that you have selected 1 point calibration in the setup menu.
1.3 Press CAL key, the meter enters calibration mode, the display shows
“
100%/CAL1
”
.
1.4 Hold the DO probe in the air, press ENTER key to confirm. Wait for the measured value to stabilize, the display shows
“END”.
Single point calibration is completed.