BANTE900P PORTABLE MULTIPARAMETER WATER QUALITY METER 

14 

 
 

MULTI-POINT CALIBRATION: 
2.1  Make sure that you have selected 2 to 5 points calibration in the setup menu. 
2.2  Repeat steps 1.2 to 1.5 above. When the first calibration point is completed, the display shows 

“CAL2”. The meter prompts you to 

continue with second point calibration. 

 

2.3  Rinse the pH electrode with distilled water. 
2.4  Dip the electrode into the next standard buffer solution (e.g. pH4.01). The meter automatically sense the current calibration 

solution and begin calibrating, 

“Calibration” indicator 

begins flashing. 

 

 

 
2.5  Wait for the measured value to stabilize, the display shows 

“CAL3”. The

 meter prompts you to continue with third point calibration. 

2.6  Repeat steps 2.3 to 2.4 

above until the display shows “END”. The meter 

automatically returns to measurement mode. Calibration is   

completed. 

 

Содержание Bante900P

Страница 1: ...Bante900P Portable Multiparameter Water Quality Meter Instruction Manual BANTE INSTRUMENTS LIMITED Technology to create the future...

Страница 2: ...lectrolyte solution membrane cap and batteries Display Bante900P portable meter is equipped with a clear and bright LCD display that used to show measured values mode indicators and message indicators...

Страница 3: ...Press the key to enter the calibration mode Press and hold the key to enter the setup menu MI Press the key to store current measured value Press key in setup mode to scroll up through menu Press key...

Страница 4: ...e the battery cover from meter s backside 2 Insert the batteries into the battery compartment note polarity 3 Replace the battery cover into its original position Installation is completed When batter...

Страница 5: ...Solutions Bante900P portable meter is packaged with three pH buffer packets required for calibration Open the pH7 00 buffer packet place the powder into a 250ml volumetric flask Pour distilled water...

Страница 6: ...or tap water for at least 15 minutes DO NOT use distilled or deionized water it will shorten the life of sensor For the Ion Selective Electrodes Soak the electrode in the diluted standard solution e...

Страница 7: ...ration Points Select the number of calibration points 1 point 2 points 3 points 4 points 5 points Resolution Sets the resolution for pH measurement 0 001pH 0 01pH 0 1pH Measurement Unit Sets the defau...

Страница 8: ...e Coefficient Select the temperature coefficient of sample solution Range 0 0 10 0 2 10 Calibration Points Select the number of calibration points 1 point 2 points 3 points 4 points 5 points Pure Wate...

Страница 9: ...grams per liter ppm Parts per million mmHg Pressure unit kPa Pressure unit GENERAL OPTIONS PARAMETER DESCRIPTION OPTIONS DESCRIPTION DEFAULT Stable Criteria Sets the stability standard for measurement...

Страница 10: ...nction allows user to restore the meter back to factory default parameters When this function is enabled all calibration values and selected system parameters will be lost or reset Enable Disable Sett...

Страница 11: ...ess ENTER key to confirm the meter returns to measurement mode Setting is completed You can easily set each default parameter as per the above method Setting the Date and Time 1 Press and hold the key...

Страница 12: ...easurement mode Setting is completed Temperature Compensation In order to get accurate measuring results you need to enable the manual or automatic temperature compensation before measurement or calib...

Страница 13: ...to set the temperature value of sample 4 Press ENTER key to confirm the meter returns to measurement mode Setting is completed In the temperature setting mode press or key once the setting value will...

Страница 14: ...ll not be accepted Make sure to calibrate the meter when attaching a new electrode or during first use DO NOT reuse calibration solution after calibration contaminants in solution will affect the cali...

Страница 15: ...bration 2 3 Rinse the pH electrode with distilled water 2 4 Dip the electrode into the next standard buffer solution e g pH4 01 The meter automatically sense the current calibration solution and begin...

Страница 16: ...tinue with second point calibration 3 5 Rinse the pH electrode with distilled water Dip the electrode into the next custom buffer solution 3 6 Press or key to increase or decrease displayed value pres...

Страница 17: ...ows LOC 2 Press or key until the display shows ELE Electrode Diagnosis 3 Press ENTER key to confirm the meter shows the last calibration date Format mm dd 4 Press key the meter shows OFS Zero point of...

Страница 18: ...ss MODE key until the meter shows indicator 2 Rinse the ORP electrode with distilled water Dip the electrode into the calibration solution Stir the sensor gently 3 Press CAL key the meter shows curren...

Страница 19: ...de 1 Press MR key in the ORP measurement mode the meter shows LOC 2 Press or key until the display shows ELE Electrode Diagnosis 3 Press ENTER key to confirm the meter shows the last calibration date...

Страница 20: ...te measurement results we recommend that you perform ion calibration and measurement at same temperature 1 Press MODE key until the meter shows indicator 2 If necessary select the concentration unit a...

Страница 21: ...sured value to stabilize the display shows 0 1 CAL2 The meter prompts you to continue with second point calibration 7 Dip the Ion Selective Electrode into the corresponding calibration solution Stir t...

Страница 22: ...in the ion measurement mode the meter shows LOC 2 Press or key until the display shows ELE Electrode Diagnosis 3 Press ENTER key to confirm the meter shows the last calibration date Format mm dd 4 Pre...

Страница 23: ...ccuracy we recommend that you perform 3 points calibration or select a standard value close to the sample value you are measuring The meter will automatically detect these conductivity standard soluti...

Страница 24: ...r returns to measurement mode automatically Calibration is completed MULTI POINT CALIBRATION 2 1 Make sure that you have selected 2 to 5 points calibration in setup menu 2 2 Repeat steps 1 2 to 1 5 ab...

Страница 25: ...ts you to continue with third point calibration 2 7 Repeat steps 2 3 to 2 5 above until the display shows END the meter returns to measurement mode automatically Calibration is completed MANUAL CALIBR...

Страница 26: ...mode 3 6 Press or key to set each digit according to previous record press ENTER key to confirm each selection until the setting values stop flashing The meter returns to measurement mode automatical...

Страница 27: ...measurement mode the meter shows LOC 2 Press or key until the display shows ELE Electrode Diagnosis 3 Press ENTER key to confirm the meter shows the last calibration date Format mm dd 4 Press key the...

Страница 28: ...saturation calibration in the air If you select 2 points calibration you need to use zero oxygen solution 100 SATURATION CALIBRATION 1 1 Press MODE key until the meter shows measurement unit 1 2 Make...

Страница 29: ...ro oxygen solution stir the probe gently 2 5 Press ENTER key to confirm Calibration indicator begins flashing 2 6 Wait for the measured value to stabilize the display shows 8 25 CAL2 The meter prompts...

Страница 30: ...linity and atmospheric pressure coefficient in the setup menu Refer to Setup Menu section 1 4 Press CAL key the meter shows current calibration point e g 8 25mg L 1 5 Dip the DO probe into the air sat...

Страница 31: ...ro oxygen solution stir the probe gently 2 5 Press ENTER key to confirm Calibration indicator begins flashing 2 6 Wait for the measured value to stabilize the display shows 8 25 CAL2 The meter prompts...

Страница 32: ...e temperature value 3 Press ENTER key to confirm Calibration is completed pH Measurement 1 Press MODE key until the meter shows indicator 2 Rinse the pH electrode thoroughly with distilled or de ionis...

Страница 33: ...shows indicator 2 Rinse the Ion Selective Electrode thoroughly with distilled or de ionised water 3 Dip the electrode into the sample solution stir the sensor gently 4 Wait for the reading to stabiliz...

Страница 34: ...ater brine and other liquids If you use meter to measure seawater or other water containing large amounts of salt please set salinity coefficient before use Some gas and steam such as chloride sulfur...

Страница 35: ...immediately freeze currently displayed value Press the key again to release the measured reading you can continue to take measurement Storing and Recalling Data from Memory The meter allows up to 500...

Страница 36: ...ER QUALITY METER 35 3 Press key the meter shows the date and time of the stored data set Format mm dd 4 Press key again the display shows the stored data 5 Press MEAS key the meter exits current mode...

Страница 37: ...System icon on the start bar the DAS splash screen appears for few seconds and then the application starts RECEIVING THE STORED DATA Connect the USB cable to meter and computer Communication port COM...

Страница 38: ...data If you need to view previous data click the Recall button the screen will show all data in the database DELETE THE DATA If you need to delete the certain data click on the data field then click...

Страница 39: ...tton the screen will show a dialog box Selecting the time interval and recording time click the OK button the measured value will automatically transfer to data sheet CREATE THE EXCEL DATA FILE When t...

Страница 40: ...f HCL Place the electrode in the solution for 10 minutes Rinse the sensor with distilled water REACTIVATING THE ELECTRODE If stored and cleaned properly the electrode should be ready for immediate use...

Страница 41: ...r paper and install the probe Troubleshooting LCD DISPLAY CAUSE CORRECTIVE ACTION Electrode dried out Soak the pH electrode in 3M KCL solution for 10 minutes Soak the Ion Selective Electrode in the di...

Страница 42: ...elective Electrode Accuracy 0 5 F S Monovalent 1 F S Divalent Resolution 0 001 0 01 0 1 1 Calibration Points 2 to 5 points Calibration Solutions 0 001 0 01 0 1 1 10 100 1000 10000ppm mol L mg L Conduc...

Страница 43: ...s 1 or 2 points Temperature Compensation 0 50 C 32 122 F Pressure Correction 60 0 112 5kPa 450 850mmHg Salinity Correction 0 50g L Saturation of Oxygen Range 0 0 200 0 Accuracy 2 0 Resolution 0 1 Temp...

Страница 44: ...cts Bread Dough Cement Cosmetics Dairy Products Education Fats Cream Field Use Fish Products Lab Flasks Low Ionic Meat Cheese Micro Samples Paint Photographic Soil Surface Test Tubes Tris Buffer Visco...

Страница 45: ...phide Addendum 4 How to prepare the conductivity standard solutions Place AR potassium chloride reagent in a 50mL beaker and dry in an oven for 3 to 5 hours at 105 C then cool to room temperature in d...

Страница 46: ...800 91 7 688 900 90 5 679 1000 89 4 671 1100 88 3 662 1200 87 2 654 1300 86 1 646 1400 85 0 638 1500 84 0 630 1600 82 9 622 1700 81 9 614 1800 80 9 607 1900 79 9 599 2000 78 9 592 Addendum 6 How to pr...

Страница 47: ...hese efforts and we expect to have little or no such materials in our product in the coming years We welcome our customer suggestions on how to speed up these efforts Warranty The warranty period for...

Отзывы: