BANTE900P PORTABLE MULTIPARAMETER WATER QUALITY METER
31
Temperature Calibration
During the measurement, if the temperature reading displayed differs from that of an accurate thermometer, you need to calibrate the
meter.
1. Press and hold the
o
C key for 3 seconds to enter temperature calibration mode, the display shows current temperature reading.
2. Press
or
key to set the temperature value.
3. Press ENTER key to confirm. Calibration is completed.
pH Measurement
1. Press MODE key until the meter shows
indicator.
2. Rinse the pH electrode thoroughly with distilled or de-ionised water.
3. Dip the electrode into the sample solution, stir the sensor gently.
4. Wait for the reading to stabilize, record the measured value on the display.
ORP Measurement
Bante900P portable meter provides two ORP measurement modes.
ABSOLUTE MILLIVOLT MODE:
Press MODE key until the display shows
measurement unit “mV”, the meter is now enter
s absolute mV measurement mode.
mV