BANTE900P PORTABLE MULTIPARAMETER WATER QUALITY METER
44
Addendum 3: Optional Ion Selective Electrodes
The meter can connect with various types of composite ion selective electrodes. The following list shows the selectable sensors.
MODEL
ION
F-US
Fluoride
CL-US
Chloride
Br-US
Bromide
Cn-US
Cyanide
Na-US
Sodium
NO3-US
Nitrate
Ca-US
Calcium
NH4-US
Ammonium
Cd-US
Cadmium
NH3-US
Ammonia
Cu-US
Cupric
I-US
Iodide
Pb-US
Lead
K-US
Potassium
Ag-US
Silver
S-US
Sulphide
Addendum 4: How to prepare the conductivity standard solutions
Place AR potassium chloride reagent in a 50mL beaker and dry in an oven for 3 to 5 hours at 105°C, then cool to room temperature in
desiccator.
Calibration Solutions
84µS/cm
Accurately weigh out 42.35mg of KCL, and dissolve in 1000mL de-ionised water
1413µS/cm
Accurately weigh out 745.9mg of KCL, and dissolve in 1000mL de-ionised water
12.88mS/cm
Accurately weigh out 7.45g of KCL, and dissolve in 1000mL de-ionised water
111.8mS/cm
Accurately weigh out 74.5g of KCL, and dissolve in 1000mL de-ionised water
Stir the solution until the solution has thoroughly mixed. Preparation is completed.