ZM400 Series Indicators Service Manual
75
6.5
Ports
PAth
Use this to choose where the file is saved, internally for
transfer by
FtP
using an Ethernet port or to a connected
uSb
flash drive.
ModE
Use this to configure the file convention. Choices are:
conStAn
This will create a single file that will be appended
to as new information is saved.
nuMbrd
This will create new file appended with a
sequential number for each transaction.
d_t
This will create a new file and the file name will be
appended with the date and time (including
seconds) for each transaction.
rESEt
Reset will clear all stored transactions and also reset the
numbered file sequence to 000001.
When the FILE > PATH is set for USB, a USB flash drive can be installed into the USB
Host port and transactions will be transferred to the flash drive as they occur. If the
flash drive is not installed, transactions will be saved onto the Flash memory in the
indicator.
The size of the print format will determine the number that can be stored but with a
typical print format the indicator should be capable of storing several thousand
transactions. The saved transactions will transfer immediately when the USB flash
drive is re-installed. The message
CoPYing
will display as each stored file is
transferred to the flash drive. If using the d_t file mode this could take a few seconds
to transfer each file. Do not remove the flash drive until the message
donE
is finally
displayed. Stored transactions will then be cleared after the transfer.
Filename Examples:
<MODEL>_<SERIAL>XXXXXX.txt
where <MODEL> is the model number, <Serial> is the serial number, XXXXXX is
either nothing, the sequential number, or the time and date.
Sequential number increments from 000001 – 999999 …. when RESET starts at
000001
Date = YYYY_MM_DD__HH_MM_SS
with double underscore between DD and HH
Examples
:
Constant:
ZM400_120450001.txt
Numbered: ZM400_120450001_000001.txt
D_T: ZM400_120450001_2012_01_31__10_31_23.txt
If transactions are being store on the indicator Flash memory, only a single Constant
or Numbered file will be created. If using the Date & Time file naming convention then
individual files will be created.
Содержание ZM401
Страница 1: ...AWT35 501334 Issue AA ZM401 405 Weight Indicators Service Manual original instructions CONFIDENTIAL...
Страница 8: ...8 ZM400 Series Indicators Service Manual...
Страница 116: ...116 ZM400 Series Indicators Service Manual 10 Print formatting...
Страница 130: ...130 ZM400 Series Indicators Service Manual...
Страница 145: ...ZM400 Series Indicators Service Manual 145 13 9 Outline dimensions stainless steel ZM4xx Stainless Steel...
Страница 146: ...146 ZM400 Series Indicators Service Manual 13 10 Outline dimensions panel mount ZM4xx Panel Mount...
Страница 150: ...150 ZM400 Series Indicators Service Manual...
Страница 154: ...154 ZM400 Series Indicators Service Manual WSTAT token 124 Z Zero calibration 41 ZM305 main PC board 142...
Страница 155: ......