43
6.2
Calibration Procedure
Pt2-Pt9
The points are numbered 2 through 9 because, internally, the
zero reference point is point 1 and the span point is point 10.
Follow the same steps as described in the SPAN procedure to
enter each linearity point.
6.2.5 Input procedure
ZEro
Use this to enter a value for the zero point.
CountS
Use this to enter a zero point using ADC counts.
mV
Use this to enter a zero point using a mV/V value.
SPan
Accept the flashing displayed span weight (
XXXX
) or key in the span
weight that corresponds with the span ADC or mV/V value.
CountS
Use this to enter a span using ADC counts.
mV
Use this to enter a span using a mV/V value.
Linearity points are cleared if a new span calibration is performed.
Input
Zero
Span
Counts
mV
Counts
mV
XXXX
The BSQ base only allows entry of Counts for Zero or Span points
The Span value is the differential value of the actual Calibration Zero and Span count
(or mV/V) values.
Содержание ZM401
Страница 1: ...AWT35 501334 Issue AA ZM401 405 Weight Indicators Service Manual original instructions CONFIDENTIAL...
Страница 8: ...8 ZM400 Series Indicators Service Manual...
Страница 116: ...116 ZM400 Series Indicators Service Manual 10 Print formatting...
Страница 130: ...130 ZM400 Series Indicators Service Manual...
Страница 145: ...ZM400 Series Indicators Service Manual 145 13 9 Outline dimensions stainless steel ZM4xx Stainless Steel...
Страница 146: ...146 ZM400 Series Indicators Service Manual 13 10 Outline dimensions panel mount ZM4xx Panel Mount...
Страница 150: ...150 ZM400 Series Indicators Service Manual...
Страница 154: ...154 ZM400 Series Indicators Service Manual WSTAT token 124 Z Zero calibration 41 ZM305 main PC board 142...
Страница 155: ......