MAN - BBR3 - REV: Original | B. Safety
Do not submerge hand and arm in extremely cold test bath. Do not attempt to pull the drain plug
when the fluid bath level is high and temperature is extremely cold. Extremely cold fluid may cause
frost bite.
Use caution when handling air hoses during operation or when performing maintenance as contents
will be under pressure.
Handle load cell with care. The load cell will be damaged if the load frame is put on it’s side with the
load nose attached, or if the load nose is over tightened.
Handle the refrigeration hose with care. The refrigeration hose may be damaged if it is twisted or
pulled, especially when the unit is cold. Do not move the refrigeration unit unless it has been turned
off for at least 8 hours and has reached room temperature.
Prevent damage to the cooling unit. Never operate the cooling unit if the ambient temperature is
higher than 25 degrees C.
The cooling unit should be placed in a well- ventilated area, on a stand at least six inches above the
floor and not on the same surface as the BBR3.
Carefully place the calibration weights on the load cell’s weight pan. If the Linear Variable Differential
Transducer (LVDT) shaft is bumped, it may become inaccurate or even permanently damaged.
Avoid damage to the Confidence Beam. Do not leave a load on the confidence beam for an extended
period of time. This may cause the beam to bend and could result in inaccurate readings for future
tests.
Before energizing the electrical power to the bending beam rheometer, turn off all power switches and
place all controls in an OFF or neutral position. Check that your power source is of the appropriate
voltage and is surge- protected. Use appropriate power adapters based upon your region.
FLAMMABLE CHEMICALS may be located within enclosure. Exposure may result in severe injury.
Refer to maintenance manual before servicing.