MAN - BBR3 - REV: Original | C. System Overview
15
Data Instruments
The Data Instruments box and Signal Conditioner Modules, along with the 3 Solid State Relays, are all located
on the RIGHT side of the BBR3 base unit as viewed from the front (see Figure C.10).
1. DATA ACQUISITION BOX - Digitizes the amplified voltages from the Load Cell and Signal Conditioner
Modules and sends that information to the PC. Also receives command signals from the BBR3 Computer
and sends those signals to the Solid State Relays to operate the stirrer motor, bath heater, and load control
solenoid valve.
2. LOAD CELL SIGNAL CONDITIONER MODULE - Receives and conditions the output of the Load Cell, and
supplies an amplified voltage output to the Data Acquisition Box.
3. LVDT SIGNAL CONDITIONER MODULE - Receives and conditions the output of the LVDT
Displacement Sensor, and supplies an amplified voltage output to the Data Acquisition Box.
4 - 6. DC VOLT SOLID STATE RELAYS - Used to control the functions of the Bath Heater (4), Stirrer Motor (5),
and Load Control Solenoid Valve (6). These modules are identical and interchangeable.
Figure C.10 - Data Instruments