MAN - BBR3 - REV: Original | F. Troubleshooting
42
F. Troubleshooting
F.1 Preface
Listed within this section are the most common troubleshooting errors that operators may encounter when using the
BBR3. Users may follow the steps provided to work through these basic errors.
Any additional issues or system errors should be brought to the attention of the Applied Test Systems Service
Department immediately by c1-724-283-1212 or emailing
DO NOT attempt to independently fix any other system errors. Any additional errors fixed independent of technical
support at Applied Test Systems could result in damage to the equipment, or injury on the part of the operator.
WARNING:
To prevent electrical shock, use extreme caution when removing covers or panels. Follow
your company’s electrical safety procedures thoroughly.
F.2 Load Shaft Stuck or Stalled During Verification or Standardization
When the load shaft seems to be stuck or stalled during verification and standardization procedures, the most likely
cause is a misaligned LVDT. To fix this issue, perform the following steps:
1. In order to determine if the LVDT is misaligned, the LVDT must be viewed from above.
• Remove the front cover on the load frame assembly.
• Shine a flashlight at the lower end of the LVDT shaft. This light reflects up the shaft, so the LVDT can be viewed
from above. See Figure F.1.
SYMPTOM
During load cell Verification and Standardization
results are out of range.
VERIFY LVDT ALIGNMENT
Perform LVDT alignment check procedure.
ALIGN LVDT
Perform LVDT alignment procedure.
ADDITIONAL ERROR
Contact ATS service department.