![ATS BBR3 Скачать руководство пользователя страница 49](http://html1.mh-extra.com/html/ats/bbr3/bbr3_instruction-manual_3005234049.webp)
MAN - BBR3 - REV: Original | G. Maintenance
44
G. Maintenance
G.1 Cleaning the BBR3
1. When cleaning the surface of the BBR3, first disconnect all power sources and place all controls in an OFF
position.
2. Use a very mild cleaning agent to wipe down the outside of the unit.
NOTE: When cleaning be careful not to allow cleaning agent to enter and contaminate the fluid bath.
G.2 Changing the BBR3 Fluid Bath
1. Verify that the BBR3 has reached room temperature
before attempting to change the bath fluid.
2. Connect the provided hose to the drain in the
underside of the bath. Place the other end of the hose
in a bucket large enough to hold the amount of fluid
within the unit.
3. Carefully remove the load frame. Do not place the load
frame on its side when the load cell is in place, as it
can be easily damaged.
4. Use personal protective gear (PPG) to reach into
the room temperature bath and pull up on the lever
attached to the BBR3 plug. Once the plug is loose
enough, pull it out.
5. Once the plug has been removed, the fluid will drain
through the hose into the bucket. After the bath has completely drained, remove the house and discard the bath
fluid per fluid manufacturer’s instructions.
Figure G.1 - BBR3 plug