MAN - BBR3 - REV: Original | C. System Overview
Mechanical Overload Protection
Standard
Test Weights
Calibrated and traceable to NIST
BBR3 Weight
Base = 83 lbs , Load frame = 15 lbs
Chiller Weight
50 lbs (23 Kg)
BBR3 Size/Dimensions
H=23.5” x W=24” x D=26.5”
Rating
IP20 Enclosure rating
Sample Supports
3mm (.118”) radius stainless steel spaced 4.00in (101.6mm) apart
LVDT Displacement Transducer
6.35mm (0.25in) calibrated range to provide 2μm resolution
throughout testing and verification range.
Data Display
Large on-screen display of load, displacement, and bath temperature
provides ease of setup and operation. Real time displacement,
loading, and temperature graphs are displayed during the test cycle
and can be re-plotted and re-scaled as needed for easy viewing.
Chiller
Included (non-CFC refrigerant)
Tank Liquid Volume
1.5 Gallons (5.5 Liters)
Operating Temperature
Ambient to -40°F (-40°C)
Temperature Measurement
Platinum RTD
Power Requirements
120 VAC, 60Hz or 230 VAC, 50/60 Hz 2 AMP power rating
Air Pressure
60 PSI inlet pressure (414 kpa) @ Class 3 Quality max particle of
5um
Environmental Conditions
The BBR3 is designed for use in an industry/laboratory setting in an indoor and dry environment. The base unit
should be placed on a clean, stable work surface with the connecting mechanical refrigeration unit nearby. The
mechanical refrigeration unit should be placed 6 inches above the floor and in an area that will not constitute a
tripping hazard.
Load Frame
The load frame is an independent three-point loading device designed to apply a load of up to 4000 mN. The load
frame may be operated in the supplied fluid bath, or it may be used in ambient conditions. The load frame is designed
with space-saving design, rigid construction, and corrosion resistance.
The load frame is constructed of stainless steel plates and durable high-strength PVC uprights that are designed
to be dimensionally stable and provide accurate force control. The load frame consists of an integral free-floating
loading shaft within an air bearing to permit specimen loads in the range of 0 to 4000 mN. The air bearing also
provides reliable and rapid loading with an accuracy of ±5 mN.